Catheter ablation for atrial fibrillation using the Ablation Index-High
power strategy. Do we have the ideal target?
Oluwaseun Adeola MD*, Asad Al Aboud MD*, Travis Richardson MD*, Gregory
Michaud MD*
*Arrhythmia Section, Division of Cardiology, Vanderbilt University
Medical Center, Nashville, TN
Corresponding Author:
Oluwaseun Adeola, MD
Vanderbilt Heart and Vascular Institute
2220 Pierce Avenue,
383 Preston Research Building,
Nashville, TN 37232, United States
Phone: 857-334-9670
Email:
oluwaseun.adeola@vumc.org
Disclosures: none
Funding: none
Conflicts of interest: none
Pulmonary vein isolation (PVI) is the cornerstone of catheter ablation
for atrial fibrillation (AF). Radiofrequency (RF) ablation has evolved
significantly over the past two decades with the evolution of 3D
electroanatomic mapping, intracardiac echocardiogram (ICE), contact
force sensing catheters and catheter irrigation. Nonetheless, the
overall freedom from AF after a single ablation is only about 70% using
conventional ablation strategies(1–3). Many AF recurrences can be
attributed to incomplete PVI due to ineffective lesion delivery at the
index ablation(4).
When performing PVI, the goal of ablation is to create a transmural
lesion while avoiding collateral damage to vital structures especially
the esophagus given its proximity to the posterior wall of the left
atrium. Energy delivered during ablation is converted to heat and causes
tissue damage by resistive and conductive heating. Resistive heating
occurs when energy delivered from the ablation catheter is converted to
heat as it passes through tissue to the ground electrode while
conductive heating is due to heat energy being passively transferred
from the hot lesion core to the adjacent tissues(5). Several variables
have been evaluated in the search for ideal ablation parameters
including power and duration of energy application, impedance fall,
contact force, catheter stability and inter-lesion distance among
others. Ablation index (AI) is a proprietary objective descriptor
calculated from contact force, power and duration of a stable catheter
position in a weighted formula. AI is independent of impedance fall,
though an analysis of 1,013 ablations by Ullah et al showed that AI had
a strong correlation with impedance fall(6).
In this issue of the Journal of Cardiovascular Electrophysiology, Chen
et al reported their experience with 122 patients who underwent an
ablation index-high power (AI-HP) strategy RF ablation for AF using 50W
power, targeting AI values of 550 on the anterior left atrium (LA), 400
on the posterior wall and inter-lesion distance (ILD) 6mm. They achieved
1st pass PVI in 96.7% of cases, mean RF time was
11.5min and total procedure time was only 55.8min. All patients had
72h-Holter monitor and trans-telephonic follow up. They reported 89.4%
arrhythmia free survival among patients with paroxysmal AF and 80.4%
among patients with persistent AF at 15-month follow up. Sixty (49%)
patients had luminal esophageal temperature (LET)
>390C out of which 3 (2.5%) had
asymptomatic endoscopic esophageal erosions/erythema. Four (3%)
patients had clinically apparent steam pops during ablation with no
adverse clinical sequela.
Despite the fact that there was no comparison group in this study, their
rate of 1st pass PVI is compelling. The paucity of
randomized controlled trials on AI guided AF ablation makes it
challenging to draw strong conclusions regarding its incremental benefit
over conventional strategies. Moreover, what are the best AI targets and
ablation power for an ideal ablation lesion? Taghji et al used strict
criteria to define lesion depth and contiguity targeting ILD ≤6mm and AI
≥400 on posterior LA and ≥550 on anterior LA defined as the CLOSE
protocol while using ablation power 25-35W(7). They reported 98%
1st pass PVI and 91% single procedure freedom from
atrial tachycardia/atrial fibrillation (AT/AF) at 12 months among 104
consecutive patients with paroxysmal AF off antiarrhythmic drug therapy.
Hussein et al showed similar success rates using the same AI targets
(400/550) but higher power settings (30-40W) in 89 patients with drug
refractory AF(8). Another study by Solimene et al showed only 10.8%
atrial arrhythmia recurrence among 156 patients with symptomatic AF when
the target AI values were 330-350 on LA posterior wall and 400-450 on
anterior wall with ablation power 25-35W(9). These studies used
different ablation power and AI targets but showed comparable acute
success rates to the AI-HP strategy, which appear to be higher than
those reported in prior randomized trials using conventional ablation
techniques(1–3,10). However, caution should be taken when interpreting
success results from small, non-randomized trials. For instance, the
AI-HP study appears to have included a population with less advanced AF
since the mean LA size and LV function were within the normal range even
though 46% of their patients had persistent atrial fibrillation.
Ablation beyond PVI was rare in this study, 17 (13.9%) underwent
cavo-tricuspid isthmus ablation and only 4 patients underwent additional
linear LA ablation.
Energy delivered is a product of power and duration of application of
that power. Thus, high-power settings allow an equivalent amount of
energy to be delivered over a shorter duration to achieve defined AI
targets or impedance falls, potentially reducing total ablation and
procedure times. Indeed, the authors report an impressive mean RF time
of 11.5mins and total procedure time of 55.8mins, essentially less than
half the time reported in the CLOSE protocol or the QDOT-FAST
trial(7,11). Some of this difference may be explained by the use of deep
sedation as opposed to general anesthesia, intracardiac echocardiogram
(ICE) was not utilized and assessment of dormant PV conduction was not
routinely performed post-ablation.
While high power short duration ablation with prespecified ablation
index targets potentially saves time and appears acutely effective,
improvement in safety endpoints compared to conventional ablation
strategies has not clearly been established. In fact, the authors report
a 3% (4/122) incidence of steam pops even though there was no
documented pericardial effusion or cardiac tamponade. Much of the energy
delivered during ablation is lost to convective cooling from catheter
irrigation and surrounding blood pool. Therefore, using standard
catheter irrigation settings, ablation using lower power needs to be
delivered over a longer duration to allow the endocardium reach target
temperatures for irreversible injury. This longer duration allows more
time for conductive heating to occur in deeper tissues. Ablation using
high power settings (50W) utilizes more resistive heating to create
lesions and overcome the endocardial sparing with standard irrigation.
This may potentially minimize the time for conductive heating to deeper
tissues to occur. However, a significant proportion of patients in this
study (49%) still had LET >390C although
no reported cases of atrio-esophageal fistula occurs and a small
fraction (2.5%) had asymptomatic esophageal erosions/erythema at
endoscopy. This would suggest that in the thin posterior LA, high power
settings will not somehow violate thermodynamic principles and
completely eliminate heating of deeper structures. At the same power,
decreasing catheter irrigation rate can allow target endocardial lesion
formation to occur over a shorter duration and potentially reduce
conductive heat injury to deeper tissues(12). It is also important to
note that monitoring esophageal temperature using single thermistors may
not accurately portray the risk of esophageal injury especially when
these thermistors are far from the ablation site.
The durability of PVI using the AI-HP strategy cannot be assessed based
on this study. Eighteen patients had recurrent AT/AF (12 AF, 6 AT), out
of which 9 underwent redo ablation procedures. 5 patients had durable
PVI, a rate comparable to prior published studies using conventional
ablation strategies(1–3). One must exercise caution in equating freedom
for AF with PVI durability since not all patients with reconnected PVs
will manifest recurrent AT/AF.
The AI-HP strategy only considers ablation energy settings during stable
catheter contact but does not include catheter irrigation. It also
neglects other important parameters like tissue impedance fall and local
electrogram attenuation post ablation which are more reflective of the
effects of ablation on the tissue. While AI-HP guided RF ablation may be
an attractive strategy for PVI that likely reduces procedure times and
probably has comparable efficacy to conventional medium power ablation
settings, the incidence of steam pops and continued observation of
esophageal heating suggests that the safety of this strategy still
requires further evaluation. Feedback from the ablated tissue may need
to be incorporated into optimized ablation energy parameters to further
improve outcomes, such that lower AI values may be acceptable if steep
impedance falls and electrogram attenuation occurs before target AI
values are reached.
References
1. Morillo CA, Verma A, Connolly SJ, Kuck KH, Nair GM, Champagne J, et
al. Radiofrequency ablation vs antiarrhythmic drugs as first-line
treatment of paroxysmal atrial fibrillation (RAAFT-2) a randomized
trial. JAMA - J Am Med Assoc [Internet]. 2014 Feb 19 [cited 2020
Feb 9];311(7):692–9. Available from:
http://www.ncbi.nlm.nih.gov/pubmed/24549549
2. Kuck K-H, Albenque J-P, Chun KRJ, Fürnkranz A, Busch M, Elvan A, et
al. Repeat Ablation for Atrial Fibrillation Recurrence Post Cryoballoon
or Radiofrequency Ablation in the FIRE AND ICE Trial. Circ Arrhythm
Electrophysiol [Internet]. 2019 May 22 [cited 2019 Nov
12];12(6):e007247. Available from:
http://www.ncbi.nlm.nih.gov/pubmed/31693319
3. Packer DL, Mark DB, Robb RA, Monahan KH, Bahnson TD, Poole JE, et al.
Effect of Catheter Ablation vs Antiarrhythmic Drug Therapy on Mortality,
Stroke, Bleeding, and Cardiac Arrest among Patients with Atrial
Fibrillation: The CABANA Randomized Clinical Trial. In: JAMA - Journal
of the American Medical Association. American Medical Association; 2019.
p. 1261–74.
4. Kuck KH, Hoffmann BA, Ernst S, Wegscheider K, Treszl A, Metzner A, et
al. Impact of complete versus incomplete circumferential lines around
the pulmonary veins during catheter ablation of paroxysmal atrial
fibrillation: Results from the gap-atrial fibrillation-German atrial
fibrillation competence network 1 trial. Circ Arrhythmia Electrophysiol
[Internet]. 2016 Jan 1 [cited 2020 Dec 30];9(1). Available from:
https://www.ahajournals.org/doi/10.1161/CIRCEP.115.003337
5. Kotadia ID, Williams SE, O’Neill M. High-power, short-duration
radiofrequency ablation for the treatment of AF. Arrhythmia
Electrophysiol Rev [Internet]. 2019 Mar 1 [cited 2020 Dec
25];8(4):265–72. Available from:
/pmc/articles/PMC7358956/?report=abstract
6. Ullah W, Hunter RJ, Finlay MC, McLean A, Dhinoja MB, Sporton S, et
al. Ablation Index and Surround Flow Catheter Irrigation:
Impedance-Based Appraisal in Clinical Ablation. JACC Clin
Electrophysiol. 2017;
7. Taghji P, El Haddad M, Phlips T, Wolf M, Knecht S, Vandekerckhove Y,
et al. Evaluation of a Strategy Aiming to Enclose the Pulmonary Veins
With Contiguous and Optimized Radiofrequency Lesions in Paroxysmal
Atrial Fibrillation: A Pilot Study. JACC Clin Electrophysiol. 2018;
8. Hussein A, Das M, Chaturvedi V, Asfour IK, Daryanani N, Morgan M, et
al. Prospective use of Ablation Index targets improves clinical outcomes
following ablation for atrial fibrillation. J Cardiovasc Electrophysiol
[Internet]. 2017 Sep 1 [cited 2020 Dec 27];28(9):1037–47.
Available from: https://pubmed.ncbi.nlm.nih.gov/28639728/
9. Solimene F, Schillaci V, Shopova G, Urraro F, Arestia A, Iuliano A,
et al. Safety and efficacy of atrial fibrillation ablation guided by
Ablation Index module. J Interv Card Electrophysiol. 2019;
10. Kuck K-H, Brugada J, Fürnkranz A, Metzner A, Ouyang F, Chun KRJ, et
al. Cryoballoon or Radiofrequency Ablation for Paroxysmal Atrial
Fibrillation. N Engl J Med [Internet]. 2016 Jun 9 [cited 2020 Jan
18];374(23):2235–45. Available from:
http://www.nejm.org/doi/10.1056/NEJMoa1602014
11. Reddy VY, Grimaldi M, De Potter T, Vijgen JM, Bulava A, Duytschaever
MF, et al. Pulmonary Vein Isolation With Very High Power, Short
Duration, Temperature-Controlled Lesions: The QDOT-FAST Trial. JACC Clin
Electrophysiol. 2019 Jul 1;5(7):778–86.
12. Kumar S, Romero J, Stevenson WG, Foley L, Caulfield R, Fujii A, et
al. Impact of Lowering Irrigation Flow Rate on Atrial Lesion Formation
in Thin Atrial Tissue: Preliminary Observations From Experimental and
Clinical Studies. JACC Clin Electrophysiol [Internet]. 2017 Oct 1
[cited 2021 Jan 4];3(10):1114–25. Available from:
https://pubmed.ncbi.nlm.nih.gov/29759494/